From Wikipedia, the free encyclopedia

A hemocytometer. The two semi-reflective rectangles are the counting chambers.
Loading a chamber
Hemocytometer grid (see table)

The hemocytometer (or haemocytometer) is a counting-chamber device originally designed and usually used for counting blood cells.[1]

The hemocytometer was invented by Louis-Charles Malassez and consists of a thick glass microscope slide with a rectangular indentation that creates a precision volume chamber. This chamber is engraved with a laser-etched grid of perpendicular lines. The device is carefully crafted so that the area bounded by the lines is known, and the depth of the chamber is also known. By observing a defined area of the grid, it is therefore possible to count the number of cells or particles in a specific volume of fluid, and thereby calculate the concentration of cells in the fluid overall. A well used type of hemocytometer is the Neubauer counting chamber.[2]

Other types of hemocytometers with different rulings are in use for different applications. Fuchs-Rosenthal rulings, commonly used for spinal fluid counting, Howard Mold rulings used for mold on food and food packaging products, McMaster Egg Slide ruling used for counting microbial eggs in fecal material, Nageotte Chamber ruling for counting low levels of white cells in white cell-reduced platelet components, Palmer Nanoplankton ruling for counting smaller plankters. Petroff-Hausser counter using Improved Neubauer rulings is used for bacteria or sperm counts, and is offered with varying chamber depths. The Sedgwick-Rafter Cell ruling in a hemocytometer is primarily designed for use in the microscopy of drinking water.

YouTube Encyclopedic

  • 1/3
    Views:
    948 261
    156 585
    500 033
  • Counting Cells with a Hemocytometer
  • Hemocytometer | calculation made easy
  • Hemocytometer - Counting of cells - Amrita University

Transcription

A hemocytometer is a device that is used for counting cells. It's a modified microscope slide, containing two identical wells, or chambers, into which a small volume of a cell suspension is pipetted. We have already removed 100 µL of our cell suspension and placed it in a micro-centrifuge tube. Dilute the suspension by adding 100 µL of Trypan blue. Trypan blue is a dye that helps us distinguish between living and dead cells. The dye passes through the membranes of dead cells so they will appear blue under a microscope. Living cells exclude and will appear mostly clear. Load both chambers by pipetting the suspension under the cover slip. Now place the hemocytometer under the microscope. Each chamber is divided into a grid pattern, consisting of 9 large squares. Each square has the same dimensions and contains 10 to the negative-fourth power mL of suspension. The rules for counting cells sometimes differ from lab-to-lab. In our lab, we count cells in the 4 large corner squares and the center square. Let's count the cells in the first square. One, Two, Three, Four, Five, Six, Seven, Eight... So what about the cells that are touching the outside boundaries of the square? In our lab we count the cells that touch the top and left boundaries, and we ignore the cells that touch the right and bottom boundaries. Nine. Ten. We need to count the number of both living and dead cells. Remember, the dead cells will appear blue. Occasionally you will see artifacts - objects or debris that appear blurry and don't have a well-defined shape. This is an example of an artifact. We won't include it in our count. Proper storage, cleaning, and handling of the hemocytometer will minimize the number of artifacts. There are 10 viable cells and 1 non-viable cell in the first square. Now, the top-right square. There are 9 viable cells and no non-viable cells. Next let's count the bottom-right square. There are 11 viable cells and no non-viable cells. And now the bottom-left square. There are 10 viable cells and 2 non-viable cells. And finally, let's count the cells in the center square. Sometimes cells will appear as clumps or small groups. It may be difficult to determine exactly how many cells are in a group. The method of counting clumps of cells differs from lab to lab, so be sure to follow the procedure in your lab. We will count this clump as 2 cells. There are 14 viable cells and no non-viable cells in the center square. The total number of viable, or living cells from all 5 squares is 54. The total number of non-viable cells is 3. Now that we have counted our cells, there are several calculations we need to perform. First, let's calculate the percentage of viable cells. Here's the formula. 54 viable cells, divided by 57Éthe total number of cellsâ gives us 0.947. Multiply by 100 and the percentage of viable cells is 94.7%. Next, let's determine the average number of cells-per-square. We counted 54 viable cells. We divide 54 by 5, because we counted in 5 squares. The average number of cells-per-square is 10.8 cells. Now let's calculate the dilution factor. The dilution equals the final volume divided by the volume of cells. Our final volume is 200 µL, because we started with 100 µL of cells and added another 100 µL of trypan blue. 200 divided by 100 is 2. Therefore the dilution factor is 2. Next we need to calculate the concentration of viable cells - the number of living cells/mL Our average count-per-square is 10.8. The dilution factor is 2. 10.8 times 2 times 10-to-the-fourth-power equals 216,000 cells/mL. We can write the concentration using scientific notation as 2.16 times 10-to-the-fifth power cells/mL. From our calculations, we now know the concentration of cells in our culture is 216,000 cells-per-milliliter and approximately 94.7 percent of those cells are viable, living cells.

Principles

The gridded area of the Improved Neubauer ruled hemocytometer consists of nine 1 x 1 mm (1 mm2) squares. These are subdivided in three directions; 0.25 x 0.25 mm (0.0625 mm2), 0.25 x 0.20 mm (0.05 mm2) and 0.20 x 0.20 mm (0.04 mm2). The central square is further subdivided into 0.05 x 0.05 mm (0.0025 mm2) squares. The raised edges of the hemocytometer hold the coverslip 0.1 mm off the marked grid, giving each square a defined volume (see figure on the right).[3]

Dimensions Area Volume at 0.1 mm depth
1 x 1 mm 1 mm2 100 nL
0.25 x 0.25 mm 0.0625 mm2 6.25 nL
0.25 x 0.20 mm 0.05 mm2 5 nL
0.20 x 0.20 mm 0.04 mm2 4 nL
0.05 x 0.05 mm 0.0025 mm2 0.25 nL

Usage

To use the hemocytometer, first make sure that the special coverslip provided with the counting chamber is properly positioned on the surface of the counting chamber. When the two glass surfaces are in proper contact Newton's rings can be observed. If so, the cell suspension is applied to the edge of the coverslip to be sucked into the void by capillary action which completely fills the chamber with the sample. The number of cells in the chamber can be determined by direct counting using a microscope, and visually distinguishable cells can be differentially counted. The number of cells in the chamber is used to calculate the concentration or density of the cells in the mixture the sample comes from. It is the number of cells in the chamber divided by the chamber's volume, which is known from the start, taking account of any dilutions and counting shortcuts:

[4]

where the volume of the diluted sample (after dilution) divided by the volume of the original mixture in the sample (before dilution) is the dilution factor. For example, if the volume of the original mixture was 20μL and it was diluted once (by adding 20μL dilutant), then the second term in parentheses is 40μL/20μL. The volume of the squares counted is the one shown in the table at the top, depending on the size (see figure on the right). The number of cells counted is the sum of all cells counted across squares in one chamber. The proportion of the cells counted applies if not all inner squares within a set square are counted (i.e., if only 4 out of the 20 in a corner square are counted, then this term will equal 0.2). When counting large squares with a volume of 100 nanoliter (nL), a multiplication by 10000 leads to the desired cell count per milliliter.

The parts of the hemocytometer (as viewed from the side) are identified.

For most applications, the four large corner squares are only used. The cells that are on or touching the top and left lines are counted, but the ones on or touching the right or bottom lines are ignored.[5]

Requirements

Empty hemocytometer grid at 100x power.

The original suspension must be mixed thoroughly before taking a sample. This ensures the sample is representative, and not just an artifact of the particular region of the original mixture it was drawn from.

An appropriate dilution of the mixture with regard to the number of cells to be counted should be used. If the sample is not diluted enough, the cells will be too crowded and difficult to count. If it is too dilute, the sample size will not be enough to make strong inferences about the concentration in the original mixture.

By performing a redundant test on a second chamber, the results can be compared. If they differ more than 2 times the counting error (square root of the count[6]), the method of taking the sample may be unreliable (e.g., the original mixture is not mixed thoroughly).

The counting chamber should be filled by capillary action after the chamber lid (special cover-slip with certified thickness and flatness) has been put in place. This avoids the risk that the cells could sediment/stick to the glass or some volume to evaporate before the cover-slip is placed on top and resulting in an overestimation of the cell concentration. Sedimentation is less of a problem with bacteria but evaporation, more prevalent in low-humidity air-conditioned laboratories, still has to be minimized.

Applications

  • Blood counts: for patients with abnormal blood cells, where automated counters don't perform well.
  • Sperm counts
  • Urine microscopy
  • Cell culture: when subculturing or recording cell growth over time.
  • Beer brewing: for the preparation of the yeast.
  • Phytoplankton cell counting
  • Cell processing for downstream analysis: accurate cell numbers are needed in many tests (PCR, flow cytometry), while some others require high cell viability.
  • Measurement of cell size: in a micrograph, the real cell size can be inferred by scaling it to the width of a hemocytometer square, which is known.[7]

References

  1. ^ Absher, Marlene (1973). "Hemocytometer Counting". Tissue Culture. pp. 395–397. doi:10.1016/B978-0-12-427150-0.50098-X. ISBN 9780124271500.
  2. ^ "Fig. 8. Views of an improved Neubauer ruling hemocytometer slide. (A) Top view..." ResearchGate. Retrieved 29 March 2018.
  3. ^ "Hemocytometer square size • Hemocytometer". April 11, 2013.
  4. ^ "Cell Counting with Neubauer Chamber - Basic Hemocytometer Usage, Óscar Bastidas" (PDF).
  5. ^ Strober W (2001). "Monitoring cell growth". In Coligan JE, Bierer BE, Margulies DH, Sherach EM, Strober W (eds.). Current Protocols in Immunology. Vol. 5. USA: John Wiley & Sons. p. A.2A.1. doi:10.1002/0471142735.ima03as21. ISBN 0471142735. PMID 18432653.
  6. ^ Shapiro, Howard (31 January 2001). "very rare events: how low can you go?".
  7. ^ "Measuring cell size with a hemocytometer". Archived from the original on 2012-06-29. Retrieved 2013-06-02.

External links

This page was last edited on 7 April 2023, at 22:36
Basis of this page is in Wikipedia. Text is available under the CC BY-SA 3.0 Unported License. Non-text media are available under their specified licenses. Wikipedia® is a registered trademark of the Wikimedia Foundation, Inc. WIKI 2 is an independent company and has no affiliation with Wikimedia Foundation.